Disposable Administration Set Specifications
(continued)
D/DI-70 and D/DI-75
Filter Size
170 Micron
Priming Volume
D/DI-70
70 ml
D/DI-75
160 ml
System 1200 Normothermic (35 – 41°C)
Fluid Delivery Range
10°C Input:
50 ml/min. to 500 ml/min.
20°C Input:
30 ml/min. to 675 ml/min.
Maximum Flow Rate
750 ml/min.
(crystalloid, 300 mmHg, 8.5 F catheter)
D/DI-100 and D/DI-150
Filter Size
170 Micron
Priming Volume
D/DI-100
65 ml
D/DI-150
155 ml
System 1200 Normothermic (35 – 41°C)
Fluid Delivery Range
10°C Input:
30 ml/min. to 650 ml/min.
20°C Input:
30 ml/min. to 950 ml/min.
Maximum Flow Rate
950 ml/min.
(crystalloid, 300 mmHg, 8.5 F catheter)
D/DI-300 and D/DI-350
Filter Size
170 Micron
Priming Volume
D/DI-300
90 ml
D/DI-350
180 ml
System 1200 Normothermic (35 – 41°C)
Fluid Delivery Range
10°C Input:
30 ml/min. to 650 ml/min.
20°C Input:
30 ml/min. to 1100 ml/min.
Maximum Flow Rate
1400 ml/min.
(crystalloid, 300 mmHg, 8.5 F catheter)
S E C T I O N 1 5 • S p e c i f i c a t i o n s
71
| Operator’s Manual
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...