Operator’s Manual
Fluid Warmer Power and Alarm Test Panel
The Power and Alarm Test Panel is located on the front of the Fluid
Warmer directly above the reservoir fill-cap. This panel contains four
pressure-sensitive buttons that are activated when pressed. Refer to the
Power Alarm Test Panel (figure on right) whose numbered call outs
correspond to a description of the button and the function it performs.
Button/Function
1
Power ON Button
The green button on the top-left of the Power and Alarm Test
Panel powers on the device. Power is applied to the on-board
compressor for the pressure chambers and the Air
Detector/Clamp. The green Automatic Operation LED on the
Fluid Warmer Display Panel illuminates when the Power ON
button is activated, with a Disposable Set in place.
2
Power OFF Button
The Power OFF is the orange button to the right of the Power
ON button on the Power and Alarm Test Panel. This button
turns power off to the unit. The green Automatic Operation
LED on the Display Panel will turn off when this button is
pressed.
3
Over Temperature Test Button
The Over Temperature Test is used to confirm the proper
operation of the Over Temperature circuitry. Testing the circuitry
requires that the Fluid Warmer is at operating temperature
(41°C). Once this is established, press and hold the button.
Then, release the button. The Over Temperature alarm
continues to function. Clear the alarm mode by turning the
device off, then back on. See Section 11,
Testing, for instruction
on performing an Over Temperature Test.
4
Fluid Warmer Alarm Signal Test Button
The Fluid Warmer Alarm Signal Test is used to confirm proper
operation of the visual and audible alarm indicators. Press and
hold this button to test circuitry. Then, release the button. The
Over Temperature alarm continues to function. Clear the alarm
mode by turning the device off, then back on.
5
Reservoir Capacity
Capacity for recirculating solution reservoir is 1.4 liters. Use
recirculating solution. Do not exceed maximum capacity.
S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s
22
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...