Operator’s Manual
Check Disposables Mode
The Check Disposables mode of the Fluid Warmer indicates a missing
or improperly installed Disposable Set.
Mode characteristics
• The Check Disposables yellow LED (a) on the Fluid Warmer’s
Display Panel is illuminated.
• An audible attention signal beeps.
• Reservoir solution circulation is stopped; fluid warming stops.
• Pressure Chambers continue to operate.
To Clear this mode
Install a disposable or check the disposable installation as follows:
• Check the position of the disposable in the #
1
Block.
• Make sure the heat exchanger is seated in the heat exchanger guide.
• Press down firmly on the #
2
Block.
• Check the position of the Gas Vent/Filter Assembly in the #4
Interlock.
Add Recirculating Solution Mode
The Add Recirculating Solution mode of the Fluid Warmer indicates
that the solution level in the recirculating solution reservoir is below its
minimum level.
Mode characteristics
• The yellow Add Recirculating Solution LED (
b
), on the Fluid
Warmer’s Display Panel, is illuminated
• An audible warning signal beeps.
• Solution circulation stops, fluid warming stops.
• Pressure Chambers continue to operate.
• Fluid flow to the patient continues.
To Clear this mode
Add recirculating solution to the Fluid Warmer’s reservoir.
S E C T I O N 8 • O p e r a t i o n
32
a
b
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...