Alarm Test Mode
The Alarm Test mode is used to test the visual and audible indicators of
the Level 1
®
H-1200 Fast Flow Fluid Warmer. This mode is entered by
pressing and holding the Alarm Test button (
d
) on the Fluid Warmer’s
Control Panel.
Mode characteristics
• All visual indicators on the Fluid Warmer’s Display Panel (e)
illuminate.
• The Fluid Warmer’s audible alarm beeps.
• When the Alarm Test button is released, the Over Temperature
LED and an audible alarm remains active.
• To clear the Over Temperature alarm, turn the Fluid Warmer OFF
and then back ON.
Over Temperature Test Mode
The Over Temperature Test mode is used to test the operation of the
Fluid Warmer’s Over Temperature Circuitry. This mode is entered by
pressing and holding the Over Temperature Test button (
f
) on the Fluid
Warmer Control Panel with the Fluid Warmer at operating temperature
(41°C).
Mode characteristics
• The red Over Temperature LED (
g
) on the Display Panel
illuminates.
• An audible warning signal beeps.
To Clear this mode
• Turn Fluid Warmer OFF.
• Turn Fluid Warmer back ON.
Temperature Display
The Temperature Display functions when the Fluid Warmer is powered
ON. Temperature is displayed in degrees Celsius.
S E C T I O N 8 • O p e r a t i o n
31
| Operator’s Manual
f
g
d
e
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...