| Operator’s Manual
Step 4
Attach the I.V. Bag Hanger
1
Slide the I.V. Bag Hanger on top of the I.V. Pole.
2
Align with tabs.
3
Press down and snap into place.
Step 5
Disinfect the Recirculating Solution Reservoir
1
Remove the fill-port plug (
a
) on the reservoir.
2
Prepare a 0.3% hydrogen peroxide/distilled water solution for
the reservoir. Mix 140 ml of 3% hydrogen peroxide solution and
1,260 ml of distilled water.
3
Fill the reservoir with 1.4 liters of 0.3% hydrogen peroxide/
distilled water solution.
4
Replace the fill-port plug.
5
Insert a Disposable Set into the Fluid Warmer.
6
Insert the power cord into a properly grounded receptacle.
7
Turn the Fluid Warmer ON. Let the solution circulate for a 30-
minute disinfection period.
8
Turn the Fluid Warmer OFF.
9
Empty the reservoir.
10
Remove the Disposable Set and discard according to established
hospital procedures.
S E C T I O N 5 • O u t o f t h e B o x — A s s e m b l y
13