| Operator’s Manual
Pressure Chamber Control Panel
The Pressure Chamber Control Panel uses a control lever to switch
from pressurized to unpressurized mode. A gauge displays pressure
levels in the Pressure Chamber.
1
Pressurize / Unpressurize Lever
This lever is used to control the pressure mode in the
Pressure Chamber.
a To Pressurize the Pressure Chamber
With Pressure Chamber door closed and latched, slide lever
to the left, all the way to the plus (+) pressurized position.
This applies 300 mmHg pressure in the Pressure
Chamber when the Fluid Warmer is turned ON.
b To Unpressurize the Pressure Chamber
To remove pressure from the Pressure Chamber, slide the
lever to the right, all the way over to the minus (–)
unpressurized position. Pressure is released on the fluid bag
in the Pressure Chamber.
2
Pressure Gauge
This gauge indicates the pressure present in the Pressure
Chamber. When the Pressurize lever is in the plus (+)
pressurized position and the Fluid Warmer is ON, this gauge
displays the operating pressure in the Pressure Chamber. The
operating pressure should be 300 mmHg.
S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s
25
a
b
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...