Operator’s Manual
Automatic Operation Air Detector/Clamp
In Automatic Operating mode, the Air Detector/Clamp monitors for
the presence of air in the Disposable Set’s Gas Vent/Filter Assembly. If
air is detected the patient line is clamped off and an audible alarm
beeps. The Air Detector/Clamp goes into Automatic Operation mode
when the Fluid Warmer is turned ON and Disposable Set is properly
installed and primed—no air is present in the Gas Vent/Filter Assembly.
Mode characteristics
• The green Automatic Operation LED (
1
) is illuminated.
• Monitoring for air in the Gas Vent/Filter Assembly is active.
• Fluid is ready to be delivered to the patient when in Automatic
Operating mode.
Check Tubing Mode
The Air Detector/Clamp’s Check Tubing mode is entered when the
patient line from the the Gas Vent/Filter Assembly is not properly
installed in the Air Detector/Clamp Slot and when the Clamp Slot door
is not closed correctly.
Mode characteristics
• The yellow Check Tubing LED (
b
) on the Air Detector/Clamp
Control Panel is illuminated.
• An audible low-priority warning signal beeps.
To Clear this mode
• Place the patient line from the Gas Vent/Filter Assembly in the
#3 Clamp Slot of the Air Detector/Clamp and close the Clamp Slot
door.
S E C T I O N 8 • O p e r a t i o n
34
b
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...