Operator’s Manual
56
S E C T I O N 1 1 • T e s t i n g
Performance Testing
Cold Start Test
Store the Fluid Warmer unit in a room where the room temperature is
approximately 21°C (70°F).
1
Put a Disposable Set in place.
2
Record the start time.
3
Turn the Power button ON.
• The green system operational indicator illuminates.
• The Air Detector/Clamp goes through the Power ON Test.
• Rapidly rising numbers will appear on the recirculating solution
temperature display.
• Within 60 seconds (±) the display should read at least 30°C.
• In 3 to 10 minutes the display should read 41°C.
The Temperature Set Point is 41.7°C (+/– 0.3°C).
Note:
If the Fluid Warmer does not pass the Cold Start Test, it should
be removed from use and returned to Smiths Medical or your local
Smiths Medical distributor.
Calibration Test
One approved way to confirm proper calibration of the recirculating
solution temperature is to use the Level 1
®
Thermal Calibration Well
(TCW) (Part Number 80-03-002). A replacement filter (F-10/30 or
F-50) must be in place to utilize the TCW.
Note:
Use of the TCW requires a digital thermometer NIST traceable
and accurate within 0.1°C. Required probe size: 0.099" OD (Outside
Diameter) maximum (0.25 cm) 0.50" — 1.50" long (1.27—3.81cm).
Alternative Calibration Test
Another approved way to confirm proper calibration of the recirculating
solution temperature is to use the Level 1
®
DSTA-40 TEMPCHECK
(Part Number 80-01-040).
Calibration Test with DSTA-40
The DSTA-40 TEMPCHECK (DSTA-40) is an electronic thermometer
used to verify the recirculating solution operating temperature of the
Fluid Warmer. The DSTA-40 uses thermistor technology to sense the
temperature of the recirculating solution temperature for the Fluid
Warmer. The recirculating solution temperature is displayed by a liquid
crystal display (LCD). The DSTA-40 is powered from the auxiliary
outlet of the Fluid Warmer. No batteries are required.
Note:
The Level 1
®
DSTA-40 TEMPCHECK is available for purchase
from Smiths Medical.
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...