5-7
The monitor has the following features after the alarm sound is
paused:
The sound of all physiological alarms and technical alarms are
switched off within the set alarm audio pause time.
The remaining alarm audio pause time is displayed in the
physiological alarm information area.
The alarm audio pause symbol is displayed in the system status
information area.
After reaching the alarm audio pause time, the monitor will
automatically exit the alarm audio pause status. You can also cancel
the alarm audio paused status by pressing the [Alarm Audio Pause]
quick key.
5.8 Switching Off Alarm Sound
If [Alarm Audio Pause] is set to [Permanent] (see 5.6.4Setting the
Alarm Audio Pause Time), press the [Alarm Audio Pause] quick key
to switch off all alarm sound.
The monitor has the following features after the alarm sound is
switched off:
The sound of all physiological alarms and technical alarms are
switched off.
The alarm audio off symbol is displayed in the system status
information area.
To exit the alarm audio off status, press the [Alarm Audio Pause]
quick key.
Warning
Pausing or switching off alarm sound may result in a hazard to the patient.
Please pay attention to the potential risks.
5.9 Latching Alarms
Physiological alarms can be set to ―latch‖ or ―non latch‖ mode.
Non latch: if you do not latch physiological alarms, their alarm
indications disappear when the alarm condition ends.
Latch: if you latch physiological alarms, all visual and audible
alarm indications remains until the alarms are reset.
To latch physiological alarms, follow the steps below:
1. Press[Main Menu]
→
[Maintenance]
→
input password
→
press
[Enter].
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...