7-13
7.6 About Arrhythmia Monitoring
Arrhythmia monitoring is intended for adult, pediatric and neonatal
patients.
7.6.1 Safety Information
Warning
If you switch off all the arrhythmias alarms, the monitor cannot give any
arrhythmias alarm when that situation occurs. This may result in a hazard
to the patient. Keep the patient under close surveillance.
Warning
Heart rate reading may be affected by cardiac arrhythmias. Do not rely
entirely on heart rate alarms when monitoring patients with arrhythmias.
Always keep these patients under close surveillance.
Caution
The priority of lethal arrhythmia alarms is always high. It cannot be
changed.
Note
Because arrhythmia detection needs a template ECG waveform as
reference which is a piece of normal ECG waveform with regular rhythm
and stable amplitude, it is necessary to re-activate the template learning
when the patient is changed, or the arrhythmia detection is incorrect. For
stronger arrhythmia detection, it is recommended to wait for a clean and
stable ECG waveform before starting ARR detection during the monitoring.
Note
During arrhythmia detection, incorrect detection might occur if the
non-ECG waveforms (e.g., square or triangle waveform) appear.
Note
Before starting the 1mV calibration signal, please turn off the arrhythmia
detection.
Note
During Arrhythmia detection, the template learning is very important. The
device requires a group of stable QRS complex waveforms to build up this
template. If the system detects Arrhythmia incorrectly, please re-activate
the template learning and obtain the correct template.
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...