13-4
Patients with severe bleeding tendency or sickle cells are
forbidden to use the NIBP measurement, otherwise local
bleeding may be caused.
Do not apply cuff on the limb where skin damage has occurred
or is expected.
When extremely low or high heart rate or serious arrhythmia of
the patient occurs. Especially auricular fibrillation will lead to
unreliable or impossible measurement.
When the patient is suffering from massive hemorrhage,
hypovolemia, shock and other conditions with rapid blood
pressure change or when the patient has too low body
temperature, the reading will not be reliable, for reduced
peripheral blood flow will lead to reduced arterial pulsation.
For patients with severe coagulation disorders, it is necessary to
determine whether to carry out automatic blood pressure
measurement according to clinical evaluation, because there is
a risk of hematoma at the friction between the limb and the cuff.
13.4 Measurement Modes
The following NIBP measurement modes are supported:
Manual - measurement on demand.
Auto: repeated measurements at the set interval. Measuring at
integral points are available.
STAT: continual rapid series of measurements over a
five-minute period, then return to the previous mode. The
monitor will not stop making measurement until the measuring
time is over 5 minutes or the operator stops it manually.
Sequence: continual automatic measurement at set duration
and interval of each stage.
In this mode, a complete measurement includes up to 7 phases (A,
B, C, D, E, F, G). You can set the measurement duration of each
cycle and the interval between two NIBP measurements.
ABPM (ambulatory blood pressure monitoring): conduct NIBP
measurements according to the set sequential mode.
Warning
―STAT‖ mode can only be used for adults.
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...