A-16
Measurement
range
30bpm
~
250bpm
20bpm
~
250bpm
Measurement
accuracy
±
2bpm or ±2%,
whichever is greater
±
3bpm
PR from IBP module
Measurement
range
20bpm
~
250bpm
Measurement
accuracy
±
3bpm
A.14 CO
2
Specifications
Compliant
standards
ISO 80601-2-55: 2018
Measurement
mode
Sidestream and mainstream
Measurement
range
0 mmHg
~
150 mmHg (0 kPa
~
20.0 kPa)
Measurement
accuracy
0 mmHg
~
40 mmHg (0 kPa
~
5.3 kPa), the error is ±2
mmHg (±0.26 kPa)
41 mmHg
~
70 mmHg (5.5 kPa
~
9.3 kPa), the error is
±5%
71 mmHg
~
100 mmHg (9.4 kPa
~
13.3 kPa), the error is
±8%
101 mmHg
~
150 mmHg (13.4 kPa
~
20.0 kPa), the error
is ±10%
Alarm limit
setting range
High limit: 0.1mmHg
~
150mmHg (0 kPa
~
20kPa)
Low limit: 0 mmHg
~
149.9mmHg (0 kPa
~
19.9kPa)
Alarm
accuracy
±
1mmHg (±0.13kPa)
Sampling rate
(sidestream)
50 ml/min fixed flow rate
Response time
and rise time
Sidestream:
response time < 9 s (KM7003 CO
2
module)
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...