11-1
Chapter 11 Monitoring Pulse Oxygen Saturation
(SpO
2
)
11.1 Introduction
Pulse oxygen saturation (SpO
2
) monitoring is a non-invasive optical
technique used to measure the amount of oxygenated hemoglobin
(HbO
2
) and pulse rate by measuring the absorption of selected
wavelengths of light. The light generated in the emitter side of the
sensor is partly absorbed when it passes through the monitored
tissue. The amount of transmitted light is detected in the detector
side of the sensor. When the pulsative part of the light signal is
examined, the amount of light absorbed by the hemoglobin is
measured and the pulse oxygen saturation can be calculated. The
monitor is calibrated to display functional oxygen saturation.
SpO
2
monitoring is intended for adult, pediatric and neonatal
patients.
11.2 Safety Information
Warning
Check the SpO
2
sensor and cable before use. Do NOT use the damaged
SpO
2
sensor.
Warning
DO NOT stare at the light of SpO
2
sensor (infrared light is invisible) when it
is switched on, for the infrared light may do harm to the eyes.
Warning
Inspect SpO
2
sensor application site every one to two hours to ensure skin
quality and correct optical alignment, and change the measuring site
periodically if necessary. If the skin quality changes, move the sensor to
another site.
Warning
SpO
2
measuring site must be examined more carefully for some special
patients. Do NOT place the SpO
2
sensor on the finger with edema or
fragile tissue.
Warning
Avoid placing the SpO
2
sensor on the same extremity with an arterial
catheter, blood pressure cuff, or intravascular infusion line, otherwise the
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...