12-1
Chapter 12 Monitoring Temperature (Temp)
12.1 Introduction
The patient’s body temperature can be measured by means of a
thermistor probe (a semiconductor whose resistance changes with
temperature). Very small amount of constant current is applied to the
temperature probe to avoid self-heating. The voltage across the
thermistor is measured, and further converted into a temperature
reading according to the temperature-resistance characteristic for a
specific type of thermistor.
Up to two temperature sites can be monitored simultaneously and
the difference (
△
T) between two measured sites is calculated.
Temperature monitoring is intended for adult, pediatric and neonatal
patients.
12.2 Safety Information
Caution
Verify that the probe detecting function works correctly before monitoring.
Remove the temperature probe cable from the temperature probe
connector, and check that the monitor can display the alarm message
[Temp 1 sensor off] or [Temp 2 sensor off] and give alarm tones correctly.
Caution
When unplugging the probe, be sure to hold the head of the connector and
pull it out.
Note
There is a range of normal body temperature. At the same time, the
temperature of different measurement sites is different. Therefore, the
readings of different sites should not be compared directly.
12.3 Temp Display
2
3
1
4
2
4
5
3
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...