A-9
rejection
capability
350ms, duration of 180ms and amplitude lower than
1.2mV, the heart rate calculation is not affected.
Minute heart rate
averaging
algorithm
The lastest 8 RR intervals are all in a certain range of
average heart beats.
Display refresh rate: 1 beat/s
Response time to
heart rate
changes
Meets the requirements of IEC 60601-2-27: Clause
201.7.9.2.9.101 b) 5).
From 80 bpm to 120 bpm: less than 8 s
From 80 bpm to 40 bpm: less than 8 s
Accuracy of heart
rate meter and its
response to
irregular rhythm
In compliance with the requirements in Clause
201.7.9.2.9.101 b) 4) of IEC 60601-2-27, the heart rate
after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy (waveform A1): 80 bpm
Slow alternating ventricular bigeminy (waveform A2):
60 bpm
Rapid alternating ventricular bigeminy (waveform A3):
120 bpm
Bidirectional systoles (waveform A4): 90 bpm
Time to alarm for
tachycardia
<
11s (applicable to all clauses of IEC 60601-2-27)
ST Segment Analysis
Measurement
range
-2.0 mV~+2.0 mV
Measurement
accuracy
-0.8 mV~0.8 mV: ±0.02 mV or ±10%, whichever is
greater.
Beyond this range: not defined.
Resolution
0.01 mV
ST alarm high
limit
(low limit + 0.2 mV)
~
2.0 mV
ST alarm low limit
-2.0 mV
~
(high limit - 0.2 mV)
QT/QTc Analysis
Measurement
range
QT: 200 ms
~
800ms
QTc: 200 ms
~
800ms
QT-HR: 15 to 150 bpm for adult, 15 to 180 bpm for
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...