5-6
3. Set [High Alarm Audio Interval], [Medium Alarm Audio Interval]
and [Low Alarm Audio Interval].
The setting range of [High Alarm Audio Interval] is 3~15s.
Default setting is 10s.
The setting range of [Medium Alarm Audio Interval] is
3~30s. Default setting is 10s.
The setting range of [Low Alarm Audio Interval] is 16~30s.
Default setting is 10s.
5.6.4 Setting the Alarm Audio Pause Time
To set the alarm audio pause time, follow the steps below:
1. Press[Main Menu]
→
[Maintenance]
→
input password
→
press
[Enter].
2. Select the [Alarm] tab.
3. Set the [Alarm Audio Pause Time].
The alarm audio pause time can be set to [1min], [2min], [3min]
or [Permanent]. Default setting is two minutes.
5.6.5 Setting the Switch of the Apnea Alarm Off
You can choose whether switching off the Apnea alarm is
permissible or not.
To do so, please refer to the following steps:
1. Press[Main Menu]
→
[Maintenance]
→
input password
→
press
[Enter].
2. Select the [Alarm] tab
→
select the [Other] tab.
3. Set the [Apnea Alarm Off].
[Disabled] (default setting): the Apnea alarm is always on.
You cannot switch it off.
[Enabled]: you can switch off the Apnea alarm.
Warning
If you switch off the Apnea (zero respiration rate) alarm, the monitor will
not issue the Apnea alarm when that situation occurs. This may result in a
hazard to the patient. Keep the patient under close surveillance.
5.7 Pausing Alarm Sound
Press the [Alarm Audio Pause] quick key to pause the current alarm
sound.
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...