12.8 Exchanging the fixture for specimen
clamps
Depending on your individual order, the instru-
ment is delivered with either the orienting or
the non-orienting fixture for specimen clamps.
Both fixtures are interchangeable.
12.8.1 Disassembling the orienting fixture for
specimen clamps (Fig. 62.1)
Activate the handwheel locking system.
Remove the knife holder.
Remove the specimen clamp from the fixture for
specimen clamps.
Detach the four screws (1) with a no. 3 Allen key
and remove the dovetail adapter
(2)
.
Unscrew the adjusting screws
(3)
and
(4)
com-
pletely.
Completely unscrew pressure piece
(5)
with a slot-
ted screw driver and remove it together with spring
(5a)
and pin
(5b
).
Release eccentric bolt
(6)
rotating counterclock-
wise.
Evenly unscrew both screws
(7)
and
(8)
, accessible
through the bores holes
(9
), and remove the orient-
ing fixture for specimen clamps
(10)
.
To reassemble, proceed in reverse order.
12.8.2 Assembling the non-orienting fixture for
specimen clamps (Fig. 62.2)
Attach the non-orienting fixture
(11)
to the speci-
men cylinder as shown in Fig. 61.2 and secure with
the four corresponding screws
(12)
(use no. 4 Allen
key).
Attach the dovetail adapter
(2)
and tighten the four
screws
(1)
with a no. 3 Allen key.
To disassemble, proceed in reverse order.