Instruction manual V3.1 02/2001
7. Setting up the instrument
230
1
3
3
5 4
Fig. 16.1
3
6
Fig. 16.2
7.1.2 Changing the voltage selector setting
Do not change the voltage selector setting,
with the instrument being connected to the
mains.
Place the tip of the smaller one of the two screw-
drivers into the small recess
(4)
of cover
(5)
and push
very carefully, using the screwdriver as a lever, to
remove the housing.
Pull out the voltage selector housing
(3)
.
Remove the voltage selector
(6)
from the housing
and reinsert it so that the appropriate value shows
in the indication window
(1)
.
Reinsert the housing
(3)
together with the fuses into
the instrument, push lightly to lock it in place.
Check again whether the correct voltage shows in
window
(1)
.