11 Maintenance
Fig. 41
5
4
3
4
11.3 Replacing the fuses
Before exchanging the fuses, switch the in-
strument off and pull the mains plug!
When exchanging fuses, only use replace-
ment fuses with the same specification as the
original ones. - Refer to chapter 4 Technical
data for the correct specifications.
The mains fuses are located in the voltage selector
housing
(3)
.
Remove the voltage selector housing
(3)
(see chap-
ter 7.1.2 Changing the voltage selector setting.
Remove the fuses
(4)
.
Insert new fuses of the same technical specifica-
tion.
The selected voltage setting shows in the small indi-
cation window
(5)
in the voltage selector housing.
Reinsert the housing together with the fuses into
the instrument, push lightly to lock it in place.
Check again whether the correct voltage shows in
window
(5)
.