Page 33 Operating Manual Leica EM FC7 01/10
5.5
Using the antistatic device EM Crion with discharge mode
For trimming:
For perfect ribbons of sections a perfect trimmed sample is mandatory.
It is recommended to use the antistatic device for trimming, because it eliminates
the sticking of the chips on the specimen and on the blade.
Set the intensity for trimming at the full voltage = all 10 indication marks are
illuminated.
For sectioning:
Set the intensity to 5 indication marks. If the sections tend to stick on the knife edge,
increase the intensity step by step until the sections begin to float in a nice ribbon
over the knife edge. If the sections begin to lift up, decrease the intensity step by
step.
The electrode will lose its power if its metal tip becomes covered with ice or
debris. If this occurs, just clean the ionizer tip with a fine brush with the
antistatic device switched off.
The electrode tip carries high voltage and must not be touched or allowed to
touch any parts of the cryo ultramicrotome while operation.
Switch off the antistatic device prior to any manipulation in the chamber.
Never heat the chamber with the electrode inside.