Page 52 Operating Manual Leica EM FC7 01/10
5.12
Finishing work with the EM FC7
Close down and bake out
Pressing the Heat button will activate the automatic bake out of the chamber.
The FC7 cryochamber will now heat up to +110 °C. The heating cycle will be
switched off automatically and the instruments will go back to room temperature
During the heating time (approximately 90 minutes) the FC7 must be connected to
the mains.
The heating cycle can be stopped and started again any time by pressing the Heat
button.
Before heating the chamber, to avoid any damage to components and knives:
x
take out all specimens.
x
take out antistatic electrode.
x
take out diamond knives.
x
unlock torque limited screw holding the cryo specimen holder. This will
prevent mechanical damage of specimen holder and locking screw due
to heat expansion.
x
unlock knife holder. This will prevent mechanical damage of the locking
screw due to heat expansion.
x
move stereomicroscope with stereo-carrier to the side. This will prevent
steam / condensing water droplets on optical components.
5.13
Maintenance of the valves
Humidity in the valves can cause freezing
and sticking of the valves during operation.
To prevent this both valves in the block can
be easily removed and cleaned with ethanol.
Remove both steel valves with a screw
driver. The 2nd steel valve is located
underneath the 1st valve.
Before inserting the valves again, check that
the ball inside the valves is free.
Summary of Contents for EM FC7
Page 1: ...Leica EM FC7 Operating Manual Version 162220032 01 2010...
Page 6: ...Page 5 Operating Manual Leica EM FC7 01 10...
Page 21: ...Page 20 Operating Manual Leica EM FC7 01 10 Remove LN2 hose and take away foam packaging...
Page 38: ...Page 37 Operating Manual Leica EM FC7 01 10 MENU User or Sample profile 1 2...