Page 50 Operating Manual Leica EM FC7 01/10
Flip over section press to the right.
Release grid from the holder by pushing on
the small indentation at the rear end with a
small black lever or a pair of forceps and
retract grid holder.
Transfer flattened grid into the transfer
container.
The transfer container can be removed using
a pair of forceps.
The ceramic surfaces of the section press should be cleaned with 100% ethyl
alcohol before use.