Page 43 Operating Manual Leica EM FC7 01/10
To prevent possible influence of the grid on
the ionizer retract the forceps by pulling back
the assembly.
Grip the assembly on the centre to retract the forceps. Due to the backlash of
the bearings the grid will be moved towards the knife edge if the assembly will
be retracted by gripping on a different area of the assembly (e.g. at the end of
the Z-movement assembly).
Avoid turning the micrometer gauges once the forceps is retracted. The
preset position would be changed, thus the knife edge can be damaged if
repositioning of the forceps is careless performed.
Start sectioning using the discharge mode of the Crion (see chapter 5.6).
When it is time to place the section ribbon on
the grid, switch off the motor of the specimen
arm and the discharge function of the Crion.
For convenience both functions can be
switched of by the footswitch simultaneously
if “Bind Motor to Crion pedals” is enabled.