To Set Up SMART Memory
1. Touch
Timebase
in the menu bar, then
SMART Mem Setup...
in the drop-down menu.
2. Touch
the
SMART Memory
tab.
3. Under
Timebase Mode
, touch the
Set Maximum Memory
or
Fixed Sample Rate
button.
Information about your choice appears below the buttons. The calculated memory length
and time per sample point appear below the scroll buttons.
4. Touch
inside
the
Time/Division
data entry field and set a time per division.
Note:
If you are currently acquiring waveforms, you will notice a change in sampling rate as you select different modes.
5. If you selected
Sequence
mode, touch inside the
Num Segments
data entry field and
enter a value using the pop-up numeric keypad. If you want to use a timeout period, touch
the
Enable Timeout
checkbox; then touch inside the
Timeout
data entry field and enter a
value.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...