![LeCroy DDA-3000A Operator'S Manual Download Page 359](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311359.webp)
X-Stream Operator’s Manual
WM-OM-E Rev I
357
3. Touch
inside
the
Code Rate
field and make a selection from the pop-up menu:
.
If you select
Custom
, also touch inside the
Custom m/n
data entry field
and enter new
m
and
n
values, using the pop-up keyboard. See PRML Encoding in "PRML
Channel Emulation" then touch
Back
to complete this setup.
4. Touch
inside
the
Adjacent Transitions (d)
field and make a selection from the pop-up
menu:
.
In general, PR4 and EPR4 use d=0; 2/3(1,7) encoded E2PR4 uses d=1.
5. Touch
inside
the
Run Length Limit (k)
field and enter a value from 0 to 99 using the
pop-up numeric keypad.
6. Touch
inside
the
Bit Cell Time
field and enter a value from 1.00 ns to 1.00 µs using the
pop-up numeric keypad. Then press the
Measure Bit Cell Time
button. See Principle of
Equalization in "PRML Channel Emulation" then touch
Back
to complete this setup.
7. Touch
inside
the
VCO synch | data
field and enter a value from 1 to 32 using the pop-up
numeric keypad.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...