![LeCroy DDA-3000A Operator'S Manual Download Page 364](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311364.webp)
362
WM-OM-E Rev I
to complete this setup.
4. Touch inside the
Trigger On Sector x after Index
data entry field and enter a value from 0
to 999 using the pop-up numeric keypad.
5. Touch inside the Sector Pulse
Source
field and select an input channel or external source
from the pop-up menu.
6. Touch inside the Sector Pulse
Polarity
field and select positive or negative polarity from
the pop-up menu.
7. Touch inside the Sector Pulse
Level
field and enter a value from 250 mV to 250 mV using
the pop-up numeric keypad.
8. Touch inside the Index
Source
field and select an input channel or external source from
the pop-up menu.
9. Touch inside the Index
Level
field and enter a value from 250 mV to 250 mV using the
pop-up numeric keypad.
Servo Gate
Servo Gate triggers on all or selected Servo Gate pulses starting after the Index mark. To set it up,
proceed as follows:
1. Press the DDA's front panel Drive Analysis button
. The disk drive dialogs
appear.
2. Touch
the
Disk Triggers
tab.
3. In the "Disk Trigger Types" menu, touch
Servo Gate
. The Servo Gate data entry fields
appear.
4. Touch
inside
the
After index wait x servo gates
field and enter a value from 0 to 999
using the pop-up numeric keypad.
5. Touch inside the
then skip x gates between triggers
field and enter a value from 0 to 999
using the pop-up numeric keypad.
6. Touch inside the Servo Gate
Source
field and select an input channel or external source
from the pop-up menu.
7. Touch inside the Servo Gate
Polarity
field and select positive or negative polarity from the
pop-up menu.
8. Touch inside the Servo Gate
Level
field and enter a value from 250 mV to 250 mV using
the pop-up numeric keypad.
9. Touch inside the Index
Source
field and select an input channel or external source from
the pop-up menu.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...