X-Stream Operator’s Manual
WM-OM-E Rev I
113
Edge Qualified and Wait: Trigger after timeout. The broken upward-pointing arrows indicate potential triggers,
while the bold ones show where the actual triggers occur.
Qualified First Trigger
Qualified First trigger is intended to be used exclusively in Sequence Mode to speed up the trigger
rate. With Qualified First trigger, a single valid trigger is sufficient to acquire a full sequence. Other
than in Sequence Mode, Qualified First is identical to the Qualified triggers.
In data storage applications, the index pulse can be defined as the qualifier signal and the servo
gate signal as the trigger source.
To Set Up an Edge Qualified Trigger
1. In the menu bar, touch
Trigger
, then
Trigger Setup...
in the drop-down menu.
2. Touch
the
Qualify
trigger button
3. Touch
inside
the
Trigger On
data entry field and select a source on which to trigger:
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...