Qualified Parameters
Some LeCroy instruments and software packages give you the capability to constrain parameter
measurements to a vertically or horizontally limited range, or to occurrences gated by a second
waveform. Furthermore, both constraints can operate together. This capability enables you to
exclude unwanted characteristics from your measurements. It is much more restrictive than
Measure Gate, which is used only to narrow the span of the waveform for analysis, along the
horizontal axis.
Range Limited Parameters
To Set Up Range Qualifiers
1. From the menu bar, select
Measure
, then
Measure Setup...
from the drop-down menu.
2. Touch
a
Px
tab to open the setup dialog.
3. Touch
inside
the
Source
field and select a source from the pop-up menu.
4. Touch
inside
the
Measure
field and select a parameter from the pop-up menu.
5. Touch
the
Accept
tab of the right-hand dialog, then touch the
Values In Range
checkbox.
Depending on whether you select a vertical or horizontal parameter, the correct units will
be automatically displayed (V, s, Hz, dB) in the
Between
and
And
fields. Or, if you select a
simple ratio parameter (such as power factor) which yields a dimensionless number, no
units will be displayed.
6. Touch
the
Find Range
button to quickly display the most recent value of the parameter
measurement. From there it is a simple matter to set the desired range.
Amplitude Parameter
(range set manually)
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...