X-Stream Operator’s Manual
WM-OM-E Rev I
183
Figure 3
Moreover, a hysteresis range may be specified to ignore any spurious transition that does not
exceed the boundaries of the hysteresis interval. In Figure 3, Delta c2d- (1, 2) measures the time
interval separating the rising edge of the clock (trigger) from the first negative transition of the data
signal. Similarly, Delta c2d+ (1, 2) measures the time interval between the trigger and the next
transition of the data signal.
Level and Slope
For several time based measurements, you can choose positive, negative, or both slopes to begin
parameter measurements. For two-input parameters, such as Dtime@level, you can specify the
slope for each input, as well as the level and type (percent or absolute).
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...