![LeCroy DDA-3000A Operator'S Manual Download Page 244](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311244.webp)
242
WM-OM-E Rev I
4.
Touch the
Summary View
to enable a line of text
that
shows concisely the status of your last waveform and keeps a running count of how many
sweeps have passed.
5.
Touch inside the
Pass If
field, and select a Boolean condition from the pop-up menu.
6.
If you want to set up the test to end after a finite number of sweeps, touch the
Stop Test
checkbox. Then touch inside the
After
data entry field and enter a value, using the pop-up
numeric keypad.
7.
Under "If", touch either the
Pass
or
Fail
button to set the actions to occur
upon your waveform's passing or failing the test.
8.
Under "Then", touch the actions you want to occur: stop test, sound alarm, print result, emit
pulse, or save the waveform. If you want to have the results printed and your scope is not
equipped with a printer, be sure that the it is connected to a local or network printer. See
Printing.
9.
If you want to save your waveform automatically, touch the
Save Setup
. This will take you
out of the current dialog and will open the "Save Waveform" dialog. See Saving and
Recalling Waveforms.
10.
Test your Pass/Fail conditions by touching the
Force Actions Once
button. Press the
Clear All
button to quickly uncheck all checkboxes if you want to change your selections.
Comparing a Single Parameter
1. Touch
Analysis
in the menu bar, then
Pass/Fail Setup...
in the drop-down menu.
2. Touch
a
Qx
tab; a setup dialog for that position will open.
3. Touch
inside
the
Source1
field and select a source from the pop-up menu.
4. Touch
inside
the
Condition
field in the main dialog and select
ParamCompare
.
5. Touch
inside
the
Compare Values
field and select
All
or
Any
from the pop-up menu
. By selecting
All
, the test will be true only if every waveform falls within the limit
that you will set. By selecting
Any
, the test will be true if just one waveform falls within the
limit.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...