![LeCroy DDA-3000A Operator'S Manual Download Page 224](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311224.webp)
222
WM-OM-E Rev I
the other hand, tells the scope to begin on the third sample, then skip the number of samples
specified by the sparsing factor (4). In this way, the sample rate is effectively reduced.
For the sparsing factor (interval), you can set a value from 1 to 1,000,000 points. For the sparsing
offset you can set a value from 0 to 999,999.
Note:
The maximum sparsing offset that can be entered for any sparsing factor equals Sparsing Factor 1.
To Set Up Waveform Sparser
1. In the menu bar, touch
Math
, then
Math setup...
in the drop-down menu.
2. Touch the tab for the function (F1 to Fx The number of math traces available depends on
the software options loaded on your scope. See specifications.) you want to assign the
Sparse operation to.
3. Touch
inside
the
Source1
field and select an input waveform.
4. Touch inside the
Operator1
field and select
Sparse
from the
Select Math Operator
menu.
5. Touch
inside
the
Sparsing factor
field and enter a value, using the pop-up keypad.
6. Touch
inside
the
Sparsing offset
field and enter a value, using the pop-up keypad.
Interpolation
Linear interpolation, which inserts a straight line between sample points, is best used to reconstruct
straight-edged signals such as square waves. (Sinx)/x interpolation, on the other hand, is suitable
for reconstructing curved or irregular waveshapes, especially when the sampling rate is 3 to 5 times
the system bandwidth. The instrument also gives you a choice of Cubic interpolation.
For each method, you can select a factor from 2 to 50 points by which to interpolate (upsample).
To Set Up Interpolation
1. In the menu bar, touch
Math
, then
Math setup...
in the drop-down menu.
2. Touch the tab for the function (F1 to Fx The number of math traces available depends on
the software options loaded on your scope. See Specifications.) you want to assign the
Interpolate operation to.
3. Touch
inside
the
Source1
field and select an input waveform.
4. Touch inside the
Operator1
field, then touch the
Filter
button in the
Select Math Operator
menu.
5. Select
Interpolate
from the Filter submenu.
6. Touch the "Interpolate" tab in the mini setup dialog to the right of the main dialog.
7. Touch
inside
the
Algorithm
field and select an interpolation type.
8. Touch
inside
the
Upsample by
Upsampling is the factor by which sampling is increased.
field and enter a value, using the pop-up numeric keypad, if you want to enter a specific
value. Otherwise, use the
Up/Down
buttons to increment the displayed value in a 1-2-5
sequence.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...