X-Stream Operator’s Manual
WM-OM-E Rev I
213
Waveform Gated Parameters
To Set Up Waveform Qualifiers
1. From the menu bar, select
Measure
, then
Measure Setup...
from the drop-down menu.
2. Touch
a
Px
tab to open the setup dialog.
3. Touch
inside
the
Source
field and select a source from the pop-up menu.
4. Touch
inside
the
Measure
field and select a parameter from the pop-up menu.
5. Touch
the
Accept
tab of the right-hand dialog, then touch the
Values Based on
Waveform State
checkbox.
6. Touch
inside
the
When Wform
field and select the gating source.
7. Touch
inside
the
State Is
field and select
High
or
Low
from the pop-up menu. Parameter
measurements on the subject waveform will only be taken when the gating waveform is in
the selected state.
8. Touch inside the
Level Type
field and select
Absolute
or
Percent
from the pop-up menu.
9. Touch
inside
the
Level
field and enter the crossing level value at which you want
measurements to begin. Alternatively, touch the
Find Level
button to automatically select
the 50% level of your gating waveform:
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...