![LeCroy DDA-3000A Operator'S Manual Download Page 408](http://html1.mh-extra.com/html/lecroy/dda-3000a/dda-3000a_operators-manual_1866311408.webp)
406
WM-OM-E Rev I
pw50+
Pulse Width 50+
Definition
The average pulse width at the 50% point between the local feature
baseline and the local feature peak.
Description
All local features between the parameter cursors for an input waveform
are identified. The local baseline is identified for each feature, and the
height between the local baseline and peak is determined. The pulse
width is measured at 50% of the peak amplitude. The average of all width
measurements is displayed as pw50+. Provided for histograms is the
average pw50+ value for each local feature between the parameter
cursors.
Parameter Settings
Selecting this parameter displays a hysteresis-dialog for setting the
hysteresis value to a specified number of vertical divisions.
res
Resolution
Definition
The ratio of the track average amplitude for a high and low frequency
waveform.
Description
res returns, as a percentage, the ratio of track average amplitude (see taa
parameter description) for a low frequency LF and high frequency HF
waveform:
res = (taa(LF) / taa(HF)) * 100%.
Parameter Settings
Selecting this parameter displays a hysteresis-dialog for setting the
hysteresis value to a specified number of vertical divisions.
taa
Track Average Amplitude
Definition
The average peak-to-trough amplitude for all local features.
Description
All local features between the parameter cursors for an input waveform
are identified. The peak-to-trough amplitude is determined for each
feature, and the average is returned as taa. Provided for histograms is the
peak-to-trough amplitude for each local feature between the parameter
cursors.
Parameter Settings
Selecting this parameter displays a hysteresis-dialog for setting the
hysteresis value to a specified number of vertical divisions.
taa-
Track Average Amplitude-
Definition
The average local baseline-to-trough amplitude for all local features.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...