96
WM-OM-E Rev I
except a combination of channels 1
and
2, or channels 3
and
4, which yield 10 GS/s. The basic rule
is to choose either channel 1 or 2 for your first input, and either channel 3 or 4 for the second input.
Refer to Acquisition Modes in the specifications for maximum sample rates.
To Combine Channels
1. In the menu bar, touch
Timebase
; the "Horizontal" setup dialog opens.
2. Under
Active Channels
, touch
4
,
2
or
Auto
. The maximum sample rate is shown
alongside each button.
SDA 11000 DBI Controls
The SDA 11000 scope’s 11 GHz bandwidth and 40 GS/s sampling rate are achieved by an
innovative LeCroy technology called Digital Bandwidth Interleaving (DBI). This technology allows
resources to be borrowed from unused channels to multiply not only sample rate but also
bandwidth.
When 11 GHz is selected, the active channels are either C2 or C3
or both, providing 40 GS/s sampling on each channel. For 6 GHz
bandwidth, all four channels can be used, providing 20 GS/s
sampling on each channel. Channels can be set to allow 11 GHz
and 6 GHz bandwidths at the same time, as shown at left.
Bandwidth
Number of
Channels
Sample Rate
11 GHz
2
40 GHz
6 GHz
4
20 GHz
SMART Memory
Note:
When the SDA 11000 is in 11 GHz mode, only Fixed Sample Rate sampling is available.
LeCroy's SMART Memory feature ensures the highest time resolution for the time window
displayed, without aliasing. SMART Memory provides these advantages:
•
Acquisition memory is automatically allocated as needed.
•
Memory size optimization: Set Maximum Memory optimizes memory to obtain highest
sampling rate, reducing the risk of aliasing. You can set a maximum memory up to 48
Mpts.
•
Fixed Sample Rate allows setting of a specific sample rate, with the scope calculating
the amount of memory needed for a timebase setting.
•
The entire acquisition is displayed on the screen.
•
High-speed compaction shows all significant features of your waveform.
You can set a maximum memory up to 48 Mpts.
Summary of Contents for DDA-3000A
Page 1: ...LECROY X STREAM OSCILLOSCOPES O P E R AT O R S M A N U A L SEPTEMBER 2006 ...
Page 18: ...16 WM OM E Rev I BLANK PAGE ...
Page 64: ...62 WM OM E Rev I icon in Channels dialog zooms one trace zooms all displayed traces ...
Page 80: ...78 WM OM E Rev I 4 Select a language 5 Click Next when the Wizard appears ...
Page 81: ...X Stream Operator s Manual WM OM E Rev I 79 6 Select Complete setup ...
Page 157: ...X Stream Operator s Manual WM OM E Rev I 155 ...
Page 170: ...168 WM OM E Rev I 5 Read the status information in line StatusDescription ...
Page 214: ...212 WM OM E Rev I Delay Parameter Find Range selected ...
Page 263: ...X Stream Operator s Manual WM OM E Rev I 261 ...
Page 284: ...282 WM OM E Rev I ...
Page 285: ...X Stream Operator s Manual WM OM E Rev I 283 Using a Surface Plot Excel Example 5 ...
Page 324: ...322 WM OM E Rev I ...