Power fuse access
8.6 Fuse Replacement
167
|
www.lakeshore.com
8.6 Fuse
Replacement
Use this procedure to remove and replace a line fuse.
To avoid potentially lethal shocks, turn off controller and disconnect it from AC power
before performing these procedures.
For continued protection against fire hazard, replace only with the same fuse type and
rating specified for the line voltage selected.
Test fuse with an ohmmeter. Do not rely on visual inspection of fuse.
1. Locate the line input assembly on the instrument rear panel. See FIGURE 8-2.
2. Turn the power switch Off (O).
3. Remove the instrument power cord.
4. With a small screwdriver, release the drawer holding the line voltage selector
and fuse.
5. Remove existing fuse(s). Replace with proper Slow-Blow (time-delay) fuse ratings
as follows:
100/120/220/240 V
3.14 A T 250 V
5 × 20 mm
6. Re-assemble the line input assembly in reverse order.
7. Verify voltage indicator in the line input assembly window.
8. Connect the instrument power cord.
9. Turn the power switch On (l).
8.7 Factory Reset
Menu
It is sometimes necessary to reset instrument parameter values or clear the contents
of curve memory. Both are stored in nonvolatile memory called NOVRAM, but they
can be cleared individually. Instrument calibration is not affected except for Room
Temperature Calibration, which should be recalibrated after parameters are set to
default values or any time the thermocouple curve is changed.
FIGURE 8-2
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...