6.6.1 Interface Commands
143
|
www.lakeshore.com
CRVDEL
Curve Delete Command
Input
CRVDEL <curve>[term]
Format
nn
<curve> Specifies a user curve to delete. Valid entries: 21–59.
Example
CRVDEL 21[term]
—deletes User Curve 21.
CRVHDR
Curve Header Command
Input
CRVHDR <curve>,<name>,<SN>,<format>,<limit value>,<coeffi-
cient>[term]
Format
nn,s[15],s[10],n,+nnn.nnn,n
<curve>
Specifies which curve to configure. Valid entries: 21–59.
<name>
Specifies curve name. Limited to 15 characters.
<SN>
Specifies the curve serial number. Limited to 10 characters.
<format>
Specifies the curve data format. Valid entries: 1 = mV/K, 2 = V/K,
3 =
)
/K, 4 = log
)
/K.
<limit value> Specifies the curve temperature limit in kelvin.
<coefficient> Specifies the curves temperature coefficient. Valid entries:
1 = negative, 2 = positive.
Remarks
Configures the user curve header. The coefficient parameter will be calculated auto-
matically based on the first 2 curve datapoints. It is included as a parameter for com-
patability with the CRVHDR? query.
Example
CRVHDR 21,DT-470,00011134,2,325.0,1[term]
—configures User Curve 21 with a
name of DT-470, serial number of 00011134, data format of volts versus kelvin, upper
temperature limit of 325 K, and negative coefficient.
CRVHDR?
Curve Header Query
Input
CRVHDR? <curve>[term]
Format
nn
<curve>
Valid entries: 1–59.
Returned
<name>,<SN>,<format>,<limit value>,<coefficient>[term]
Format
s[15],s[10],n,+nnn.nnn,n (refer to command for description)
CRVPT
Curve Data Point Command
Input
CRVPT <curve>,<index>,<units value>,<temp value>[term]
Format
nn,nnn,±nnnnnn
<curve>
Specifies which curve to configure. Valid entries: 21–59.
<index>
Specifies the points index in the curve. Valid entries: 1–200.
<units value>Specifies sensor units for this point to 6 digits.
<temp value>Specifies the corresponding temperature in kelvin for this
point to 6 digits.
Remarks
Configures a user curve data point.
Example
CRVPT 21,2,0.10191,470.000,N[term]
—sets User Curve 21 second data point to
0.10191 sensor units and 470.000 K.
CRVPT?
Curve Data Point Query
Input
CRVPT? <curve>,<index>[term]
Format
nn,nnn
<curve>
Specifies which curve to query: 1–59.
<index>
Specifies the points index in the curve: 1–200.
Returned
<units value>,<temp value>[term]
Format
±nnnnnn (refer to command for description)
Remarks
Returns a standard or user curve data point.
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...