150
c
HAPTER
6:
Computer Interface Operation
Model 350 Temperature Controller
MDAT?
Minimum/Maximum Data Query
Input
MDAT? <input>[term]
Format
a
<input>
Specifies which input to query: A - D
(D1 - D5 for 3062 option).
Returned
<min value>,<max value>[term]
Format
±nnnnnn,±nnnnnn
Remarks
Returns the minimum and maximum input data. Also see the RDGST? command.
MNMXRST
Minimum and Maximum Function Reset Command
Input
MNMXRST[term]
Remarks
Resets the minimum and maximum data for all inputs.
MODE
Remote Interface Mode Command
Input
MODE <mode>[term]
Format
n
<mode>
0 = local, 1 = remote, 2 = remote with local lockout.
Example
MODE 2[term]
—places the Model 350 into remote mode with local lockout.
MODE?
Remote Interface Mode Query
Input
MODE?[term]
Returned
<mode>[term]
Format
n (refer to command for description)
MOUT
Manual Output Command
Input
MOUT <output>,<value>[term]
Format
n,
+
nnnnn[term]
<output>
Specifies output to configure: 1–4.
<value>
Specifies value for manual output.
Example
MOUT 1,22.45[term]—
Output 1 manual output is 22.45%.
Remarks
Manual output only applies to outputs in Closed Loop PID, Zone, or Open Loop modes.
MOUT?
Manual Output Query
Input
MOUT? <output>[term]
Format
n
<output>
Specifies which output to query: 1 - 4.
Returned
<value>
Format
+
nnnnn[term] (refer to command for description)
NET
Network Settings Command
Input
NET <DHCP>,<AUTO IP>,<IP>,<Sub Mask>,<Gateway>,
<Pri DNS>,<Sec DNS>,<Pref Host>,<Pref Domain>,
<Description>[term]
Format
n,n,dd,dd,dd,dd,dd,s[15],s[64],s[32],
<DHCP>
0 = DHCP off, 1=DHCP on.
<AUTO IP>
0 = Dynamically configured link-local addressing (Auto IP)
off, 1 = On
<IP>
IP address for static configuration.
<Sub Mask>
Subnet mask for static configuration.
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...