4.3.2 Display Contrast
65
|
www.lakeshore.com
4.3.2 Display Contrast
The front panel LCD display contrast can be adjusted for optimal viewing. The default
value should work well in most standard room temperature environments, but devia-
tions from room temperature, and extreme viewing angles can cause the display con-
trast to require adjustment for optimal viewing.
Menu Navigation:
Display Setup
Q
Display Contrast
Q
(1 to 32)
Default: 28
Interface Command:
BRIGT
4.4 Input Setup
The Model 350 supports a variety of temperature sensors manufactured by
Lake Shore and other manufacturers. An appropriate sensor type must be selected for
each input. If the exact sensor model is not shown, use the sensor input performance
chart in TABLE 4-7 to choose an input type with similar range and excitation. For
additional details on sensors, refer to the Lake Shore Temperature Measurement and
Control Catalog or visit our website at www.lakeshore.com.
Any unused input should be set to disabled.
Sensor input types
Sensor Type
Display
Message
Input Range
Excitation
Coefficient
Curve
Format
Lake Shore Sensors*
Negative Temperature
Coefficient (NTC) RTD
NTC RTD
(Cernox™)
0
)
to 300 k
)
(10 Ranges)
10 nA to 1 mA
(decade steps in
power, autorange
maintains <10 mV
or <1mV depending
on sensor exci-
tation setting;
Negative
log
)
/K
Cernox™, Carbon Glass,
Germanium, Rox™,
and Thermox™
Platinum RTD,
Rhodium-Iron RTD
(Option 3062 only)
PTC RTD
(Platinum)
0
)
to 10 k
)
(7 ranges)
100 nA to 1 mA
(decade steps in
power, autorange
maintains <10 mV
Positive
)
/K
PT-100 Series Platinum,
RF-800 Rhodium-Iron,
RF-100 Rhodium-Iron
Thermocouple
(Option 3060 only)
Thermocouple
±50 mV
NA
Positive
mV/K
Chromel-AuFe (0.07%),
Type E (Chromel-Constantan),
Type K (Chromel-Alumel),
Type T (Copper-Constantan)
Capacitance
(Option 3061 only)
Capacitance
0 nF to 150 nF (2
ranges
3.496 kHz;
1 mA/10 mA square
Positive or
negative
N/A
CS-501
Silicon Diode
(Option 3062 only)**
Diode
0 V–2.5 V
10 µA, 1 mA
Negative
V/K
DT-400 Series, DT-500 Series,
DT-600 Series
Gallium Aluminum
Arsenide Diode (Option
3062 only)
Diode
0 V– 10 V
10 µA, 1 mA
Negative
V/K
TG-120 Series
*Refer to the Lake Shore Temperature Measurement and Control Catalog for details on Lake Shore temperature sensors.
**This option will be available soon
Menu Navigation:
Input Setup
Q
Input
(A, B, C, or D)
Q
Sensor Type
Q
(Disabled, PTC RTD [Platinum], NTC
RTD [Cernox], Thermocouple, Capacitance)
Default: NTC RTD
Interface Command:
INTYPE
TABLE 4-7
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...