8.7.2 Product Information
169
|
www.lakeshore.com
8.7.2 Product
Information
Product information for your instrument is also found in the Factory Reset menu.The
following information is provided:
D
Firmware version
D
Firmware date
D
Serial number
D
Option card type
D
Option card serial number
D
Ethernet version
8.8 Error
Messages
The following are error messages that may be displayed by the Model 350 during
operation.
Error messages
Message
Description
DISABL
Input is disabled. Refer to section 4.4.
NOCURV
Input has no curve.Refer to section 4.4.8.
S.OVER
Input is at or over full-scale sensor units.
S.UNDER
Input is at or under negative full-scale sensor units.
T.OVER
Input at or over the high end of the curve.
T.UNDER
Input at or under the low end of the curve.
Cannot Communicate with
Input uP
The main microprocessor has lost communication with the sensor
input microprocessor.
NOVRAM Corrupt
Invalid data or contents in NOVRAM–when this message appears, options are provided
for resetting the instrument to default values, and for clearing all user curve locations
(21– 59). To perform the reset, set the desired parameters to “Yes”, then choose the
“Execute” option.
A temperature limit has
been exceeded
The temperature reading on a sensor input has exceeded the Temperature Limit set-
ting. A detailed message will follow, which includes a reference to which sensor input's
temperature limit has been exceeded.
*** Keypad Locked ***
An attempt has been made to change a parameter while the keypad is locked.
Refer to section 4.7.
*** Heater Short Circuit
Detected ***
A short circuit condition has been observed on heater output 1. A detailed message will
follow, which includes a reference to which output caused the condition. The output
will be turned off when this occurs.
*** Heater Open Circuit
Detected ***
An open circuit condition has been observed on heater output 1. A detailed message will
follow, which includes a reference to which output caused the condition. The output
will be turned off when this occurs.
***Heater Compliance***
TBD
*** Invalid Calibration ***
The calibration memory is either corrupt, or is at the default, uncalibrated state. This
message appears when the Model 350 is first powered on. To clear the message, and
continue with instrument start-up, press the Escape and Enter keys simultaneously.
*** Invalid Option Card
Calibration ***
The installed option card calibration memory is either corrupt, or is at the default,
uncalibrated state. This message appears when the Model 350 is first powered on. To
clear the message, and continue with instrument start-up, press the
Escape
and
Enter
keys simultaneously.
*** Firmware Update in
Progress ***
This indicates that the Model 350 is in firmware update mode.
8.9 Calibration
Procedure
Instrument calibration can be obtained through Lake Shore Service. Refer to
section 8.14 for technical inquiries and contact information.
TABLE 8-2
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...