146
c
HAPTER
6:
Computer Interface Operation
Model 350 Temperature Controller
HTRSET
Heater Setup Command
Input
HTRSET <output>,<heater resistance>,<max current>,<max user
current>,<current/power>[term]
Format
n,n,n,+n.nnn,n
<output>
Specifies which heater output to configure: 1 or 2.
<htr resistance>
Heater Resistance Setting (output 1 only):
1 = 25
)
, 2 = 50
)
.
<max current>
Specifies the maximum heater output current
(output 1 only):
0 = User Specified, 1 = 0.707 A, 2 = 1 A, 3 = 1.141 A, 4 = 2 A
<max user current>
Specifies the maximum heater output current if
max current is set to User Specified (output 1 only).
<current/power>
Specifies whether the heater output displays in current or
power. Valid entries: 1 = current, 2 = power.
Example
HTRSET 1,1,2,0,1[term]
—Heater output 1 will use the 25
)
heater setting, has a
maximum current of 1 A, the maximum user current is set to 0 A because it is not
going to be used since a discrete value has been chosen, and the heater output will be
displayed in units of current .
HTRSET?
Heater Setup Query
Input
HTRSET? <output>[term]
Format
n
<output>
Specifies which heater output to query: 1 or 2.
Returned
<htr resistance>,<max current>,<max user current>,<current/power>[term]
Format
n,n,+n.nnn,n
HTRST?
Heater Status Query
Input
HTRST? <output>[term]
Format
n
<output>
Specifies which heater output to query: 1 or 2.
Returned
<error code>[term]
Format
n
<error code> Heater error code: 0 = no error, 1 = heater open load, 2 = heater short for
output 1, or heater compliance for output 2.
Remarks
Error condition is cleared upon querying the heater status, except for the heater com-
pliance error for output 2 which does not latch querying the heater status, will also
clear the front panel error message for heater open or heater short error messages.
IEEE
IEEE-488 Interface Parameter Command
Input
IEEE <address>[term]
Format
nn
<address>
Specifies the IEEE address: 1–30. (Address 0 and 31 are reserved.)
Example
IEEE 4[term]
—after receipt of the current terminator, the instrument responds to
address 4.
IEEE?
IEEE-488 Interface Parameter Query
Input
IEEE?[term]
Returned
<address>[term]
Format
nn (refer to command for description)
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...