17
Retrieve the four values indicated in the junction box (on the
sticker):
- Value 1
- Value 2
- Measurement 1
- Measurement 2
Enter the maintenance menu (see
2.2.2).
Select function F3 by pressing or
Press to validate.
The message “
CALIB
” appears briefly.
Select the unit of weight - “
SI
” for kilograms or
“
LB
” for pounds - by pressing or
Validate by pressing
The message “
CAL 1
” is displayed for one second.
The value 1 is displayed “
XXXXX
”.
Display factory value 1 by pressing or
and then press to validate.
The message “
ACO 1
” is displayed for one second.
Measurement 1 “
XXXXX
” is displayed.
Display factory measurement 1 by pressing
or
and then press to validate.
F 4
C A L I B
S I
L b
C A L 1
X X X X X
A C O 1
X X X X X
2.2.6 CALIBRATING THE UNIT AT FOUR POINTS WITH FACTORY VALUES “F4”