14
Enter the maintenance menu.
Select function F2 by pressing or
Validate by pressing
The message “
CALIB
” appears briefly.
Select the unit of weight - “
SI
” for kilograms or
“
LB
” for pounds - by pressing or
Validate by pressing
The message “
CAL 1
” is displayed for one second.
A value “
0
” is displayed. If this value differs, reset to zero by
pressing or
Check that nothing is resting on the machine.
Press
The message “
AC0 1
” is displayed and a dot scrolls across
the bottom of the screen.
(The weighing unit is calculating measurement 1. This
operation takes around 30 seconds).
The message “
CAL 2
” is displayed for one second.
A value “
X.XXX
” is displayed.
Display the value of the standard weight (e.g. 2327) by
pressing or and
F2
S I
C A L 1
0.
A C O 1
C A L I B
L b
2.2.5 CALIBRATING THE UNIT AT TWO POINTS WITH A STANDARD WEIGHT “F2”
X.X X X
2 3 2 7.
C A L 2
Summary of Contents for IND 360P