13
Enter the maintenance menu (see
2.2.2).
Select function F3 by pressing or
Disconnect the sensors in the junction box
Press to validate.
The message “
CELLS
” is displayed indicating a sensor fault.
Press
The message “
CELL
” is displayed briefly then the message
“
BAD
” appears.
Reconnect the first sensor and wait two seconds. If the message “
Good
” is displayed, no fault is
detected. If the message “
bad
” continues to be displayed, the sensor is faulty (see the section on
connection, broken wire or faulty sensor problems).
Disconnect the first sensor, connect the second and wait for the “
Good
” or “
bad
” diagnosis.
Repeat for each sensor. When the message “
Good
” is displayed for the final sensor, press
The sensor voltage is displayed “
X.XXX
” (in V).
Note the voltage on the test sheet.
Connect each sensor again one by one to display the
corresponding voltage.
Reconnect all sensors.
The total voltage is displayed “
X.XXX
”.
Press
to exit function F3.
The message “
F1
” is displayed.
You can now choose another function by pressing or
or exit the maintenance menu by pressing
- - - - -
F 3
C E L L S
C E L L
B A D
F 1
2.2.4 TESTING THE SENSORS AND SENSOR VOLTAGE DISPLAYS “F3”
X.X X X