INSTALLATION
3
35
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
3.4.3 Step 3: setting up the units
Connect the cables of the temperature sensors or temperature transmitters with the I/O
terminal box. Connect the I/O terminal box with the converter.
•
Plug the connector
2
of the I/O box
3
in the connector
1
of the converter. Then use the
connector
4
to setup the wanted I/O, as shown in the connection diagrams on the next pages.
3.4.4 Step 4: set up I/O input for energy measurement
Go to menu nr. 2.4.5 via "Measurement
Setup
I/O"
If the recommended KROHNE energy set is used, only check the installed values of the "Current
inputs". If another energy set is used, teach in the required values. The KROHNE temperature
range is 4 mA = 0
°
C / 32
°
F and 20 mA = 120
°
C / 248
°
F. Extended range min and max is for an
alarm function. A value lower than the given minimum value and a value higher than the given
maximum will generate a signal that will be logged in the event logger.
3.4.5 Step 5: Set up process input for energy measurement
Go to menu 2.4.2.3.1 via "Measurement
Setup
Process input
Heat
Function off
"
Figure 3-19: Optional I/O box
1
I/O connector on converter
2
Connector of I/O box
3
I/O box with screw terminals for 1 mm
2
wiring
4
Screw terminal pin 1...15
2.4.5
Current output
Current output
Current output
Current output
Current inputs
Status output / limit switch
Frequency / pulse output
Simulation
13-04-2010 14:11:09 13 MB free
2.4.2.3.1
Heating
Cooling
Off
Off
Off
Off
13-04-2010 14:11:09 13 MB free