3
INSTALLATION
22
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
V mode;1 rail (DN15...250)
Press "
"
Read from the display of the signal converter what rail configuration is required and what the
correct sensor distance is.
Please follow up the instructions for the application.
One rail with 2 transducers in V mode: Put the first transducer at position "0", the other one at
the recommended distance, shown on the screen (in this example 34.30 mm).
•
Unlock the transducer by turning the locking knob
2
counter clockwise.
•
Slide the transducer
1
to the new position
3
.
•
Lock the transducer by turning the locking knob
2
clockwise.
V mode; 2 rails (DN200...750)
1.8
Advised sensor position
34.30 mm
Signal quality
78%
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
13-04-2010 14:11:09 13 MB free
INFORMATION!
For V mode with 2 rails you must install the DOWN rail in line with UP rail. It is easier to install
than the Z mode, but you need more free pipe length.