4
ELECTRICAL CONNECTIONS
46
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
Pulse/frequency output passive
•
U
ext
≤
32 VDC
•
f
max
in operating menu set to f
max
≤
100 Hz:
I
≤
100 mA
open:
I
≤
0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I
≤
10 mA
U
0, max
= 2 V at I
≤
100 mA
•
f
max
in the operating menu set to 100 Hz < f
max
≤
10 kHz:
I
≤
20 mA
open:
I
≤
0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 1.5 V at I
≤
1 mA
U
0, max
= 2.5 V at I
≤
10 mA
U
0, max
= 5.0 V at I
≤
20 mA
•
If the following maximum load resistance R
L, max
is exceeded, the load resistance R
L
must be
reduced accordingly by parallel connection of R:
f
≤
100 Hz: R
L, max
= 47 k
Ω
f
≤
1 kHz: R
L, max
= 10 k
Ω
f
≤
10 kHz: R
L, max
= 1 k
Ω
•
The minimum load resistance R
L, min
is calculated as follows:
R
L, min
= (U
ext
- U
0
) / I
max
•
Can also be set as a status output; for the electrical connection, see status output connection
diagram.
Pulse/frequency output active
•
U
nom
= 15 VDC
•
f
max
in operating menu set to f
max
≤
100 Hz:
I
≤
20 mA
open:
I
≤
0.05 mA
closed:
U
0, nom
= 15 V at I = 20 mA
•
f
max
in the operating menu set to 100 Hz < f
max
≤
10 kHz:
I
≤
20 mA
open:
I
≤
0.05 mA
closed:
U
0, nom
= 13.5 V at I
≤
1 mA
U
0, nom
= 12.5 V at I
≤
10 mA
U
0, nom
= 9.0 V at I
≤
20 mA
INFORMATION!
Pulse / frequency output
•
Any connection polarity.
•
Galvanically isolated.