Specific for sensors
WARNING!
•
Be careful when locking the rail back onto the mounting units as your fingers may get stuck
between rail and pipe it is mounted on. This may cause injury.
•
Be careful when mounting the fixation units using the metal strap. The edge of the strap may
cause injury.
CAUTION!
•
Do not bend the metal mounting strap. This may cause improper mounting of the fixation
units of the sensor rails.
•
Protect the pipe contact side of the transducer. Scratches or other damages may have a
negative impact on its proper functioning.
•
Before fitting the transducer to the transducer knob in the sensor rail, check the connection
groove of the transducer cover for damages or dirt. Clean or replace when dirty or damaged.
•
Check sensor cabling at regular intervals for damages and wear, as this may cause improper
functioning. Replace when necessary.
•
Check the sensor rail sliding area regularly for dirt or other pollution or excess coupling fat,
that may cause improper functioning.
INFORMATION!
•
Check the presence of sufficient grease on the transducer pipe contact side in case of
acoustic signal failure.
•
Excess of coupling fat may be removed from the sensor rails and transducers with a dry piece
of cloth. Coupling fat on the converter housing may be removed using soapy water.
CAUTION!
The device should be protected from corrosive chemicals or gases and dust/particles
accumulation.