2
INSTALLATION
14
OPTISONIC 6300
www.krohne.com
06/2019 - 4005857501 - QS OPTISONIC 6300 v2 R01 en
2.8.2 Long horizontal pipes
•
Install on slightly ascending pipe section.
•
If not possible, ensure adequate velocity to prevent air, gas or vapour from collecting in the
upper part.
•
In partially filled pipes, the clamp-on flowmeter will report incorrect or no flow rates.
2.8.3 Bends in 2 or 3 dimensions
Figure 2-10: Long horizontal pipes
INFORMATION!
2 dimensional bends occur in a vertical or
or
or
or horizontal plane (X/Y) only, while 3 dimensional bends
occur in both vertical and
and
and
and horizontal plane (X/Y/Z).
Figure 3-3: 2 and/or 3 dimensional bends upstream of the flowmeter
1
2 dimensions = X/Y
2
3 dimensions = X/Y/Z
for 2 path using bends in 2 dimensions:
t
10 DN; when having bends in 3 dimensions:
t
15 DN
for 1 path using bends in 2 dimensions:
t
20 DN; when having bends in 3 dimensions:
t
25 DN