4
START-UP
46
OPTISONIC 6300
www.krohne.com
06/2019 - 4005857501 - QS OPTISONIC 6300 v2 R01 en
4.3 Start measurement of large version
Before installation
Before installation
Before installation
Before installation
•
Power up the converter (do not mount and/or connect the rails yet)
•
Fill in menu X1...X5 refer to
General instructions for programming
on page 44
Select "1 path" initially in X4
•
X7.1: Check the reading with the sensor code (Ta/Tb) on rail
•
X7.2.1: Check the reading with the calibration number on the nameplate
•
X7.2.2: Check the factory preset number of traverses (default: 1 for Z-mode)
•
X7.2.3: Read the advised mounting distance. Write it down, you need it later. The installation
menu can be closed, continue with mechanical and electrical installation.
i
Mounting distance
Mounting distance
Mounting distance
Mounting distance
The advised mounting distance is needed when continuing with the configuration.
Proceed with the mechanical installation of the rails: refer to
Mechanical installation of
large version
on page 24.
After the mechanical installation of the rails, continue with the standard set up (configuration)
refer to
Start measurement (standard setup)
on page 45.
Figure 4-2: Procedure for installation of large version
1
Enter the values for the installation menu, X1...X7.2.8
2
Read the advised mounting distance in menu X7.2.3
3
Choose for Z-mode (default) or
4
Choose for V-mode