2.10.3 Turning the display of the field housing version
The display of the field housing version can be turned in 90
°
increments
1
Unscrew the cover from the display and operation control unit.
2
Using a suitable tool, pull out the two metal puller devices to the left and right of the display.
3
Pull out the display between the two metal puller devices and rotate it to the required position.
4
Slide the display and then the metal puller devices back into the housing.
5
Re-fit the cover and tighten it by hand.
Figure 2-29: Turning the display of the field housing version
CAUTION!
The ribbon cable of the display must not be folded or twisted repeatedly.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin-
free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.